top of page
Safety_Tester_new.png

SAFETY TESTER

  • Define your own test procedures

  • Windows user interface provides an easy-to-use solution

  • Data input via touch screen, keyboard or barcode scanner

  • Visual inspection, electrical safety and functional test user definable

For more than 30 years, the "Bender tester" has been a well-known term for quality and longevity in the area of fully automated electrical safety testers. The product line UNIMET® stands for standard-compliant tests and measurements, easy operation, wide interface variety and high-quality service.

​

UNIMET® provides safety for medical electrical devices used for mobile service visits and for manufacturers of medical equipment.

FEATURES

  • The Windows user interface provides an easy-to-use solution

  • Data exchange and memory via Control Center

  • Automatic, semi-automatic and manual test sequences

  • Data input via touch screen, keyboard or barcode scanner

  • Visual inspection, electrical safety and functional test user definable

  • Test sequences user-definable

  • Data memory > 10.000 data records

  • Filter function for fast data selection

  • Management of test dates

  • Multitenancy

  • Catalogue system

  • Test probe with two switching contacts for semi-automatic testing of conductive parts not connected to PE

​

Wide variety of interfaces

  • RS-232 interface

  • USB port for printer, scanner, keyboard and USB stick connection

  • PS/2 port for scanner and keyboard

  • Network connection via Ethernet (optional)

​

Standard accessories

  • Carrier bag of robust design and universal carrying concept

  • Measuring lead with test probe

  • Measuring lead with test clip

  • Null-modem cable

  • VK701-7 for testing non-heating devices

  • Data transfer software UNIMET® Control Center

​

Service

  • 36-month calibration interval

  • 36-month warranty

  • Software updates to be always up-to-date

APPLICATIONS

  • IEC 60601-1, 3rd Edition (*)

  • IEC 62353:2014

  • DIN EN 62353 (VDE 0751-1):2015-10

  • IEC 61010-1:2011-07 1)

  • ÖVE/ÖNORM EN 62353:2009-01

  • DIN VDE 0701-0702:2008-06

  • ÖVE E8701-1:03-01
    (*) These standards can be activated as a license in the basic device.

​

  • Standard-compliant measurements

  • PE conductor resistance

  • Insulation resistance

  • Earth leakage current

  • Protective conductor current

  • Equipment leakage current

  • Touch current

  • Touch voltage

  • Direct measurement method

  • Differential measurement method

  • Equipment leakage current-alternative method

  • Applied part leakage current-alternative method

  • Applied part leakage current AC and DC

  • Patient auxiliary current AC and DC

  • Applied part leakage current – mains voltage on F-Type